Category:Microbiological Sample Preparation
Model:KN580
Overview:The Kjeldahl method is to convert nitrogen-containing organic matter into inorganic nitrogen ammonium sulfate for detection, and multiply the measured value of nitrogen content by a certain coefficient to obtain the protein content. KN580 Kjeldahl Nitrogen Analyzer is equipped with distillate temperature detection function to avoid ammonia volatilization caused by excessive distillate temperature
KN580 Kjeldahl Nitrogen Analyzer adopts 16-bit three-base chromaticity color sensor, high-precision fuzzy search titration end-point determination technology, combined with unique titration system structure to ensure accurate titration; the self-developed alkali pump anti-crystallization technology solves the problem of alkali pump crystallization The reliability of the instrument is improved; the modular structure design and multiple safety protections ensure the stability, reliability and accuracy of the product.
Automatic Detection
Acid addition, alkali addition, distillation, titration, result calculation and printout are fully automatic, and automated sample testing
Sample Explosion Proof
Unique sample dilution and alkali addition methods can be set arbitrarily according to different user samples to prevent the occurrence of sample bumping and improve the measurement accuracy
Alkali Pump Anti-crystallization
The self-developed alkali pump anti-crystallization technology solves the problem of alkali pump crystallization and improves the reliability of the instrument
Steam Adjustable
The steam flow can be automatically adjusted to better achieve the distillation effect and ensure the accuracy of the experiment
Dual Mode Titration
Using dual-mode titration, the combination of thickness and fine titration will ensure the high accuracy of the results; the titration method conforms to AOAC, EPA, Din, ISO and other certification methods, and the titration adopts the officially recognized color titration method
End Point Judgment
End-point color user fine-tuning technology and end-point color learning technology ensure the adaptability of accurate testing and show more humane care
Titration Vessel Empty
After the sample is distilled, the instrument automatically completes the emptying of the residual waste liquid in the sample tube, and the titration cup is automatically drained and cleaned.
Safe Protection
Multiple intelligent protections such as digestive tube presence detection, safety door, condensate detection, steam generator overheating, overpressure protection, liquid level detection, etc. ensure the safety of instruments and operators
1. Measurement range: 0.1-250mgN (milligram nitrogen)
2. Measurement speed: 3-8min/sample
3. Titration accuracy: 1μl/step
4. Reproducibility: 0.5% relative error
5. Recovery rate: above 99.5%
6. Automatic titration: judging method of color end point according to AOAC standard
7. Data storage: >100,000 sets of data
8. Program storage: >10000 programs
9. Distillation time length: 1-9999min;
10. Steam power level: 1%-100%
11. Determination of sample volume: solid ≤ 5g liquid ≤ 25ml;
12. Circulating water flow: 1.5 liters/min