Category:Chromatography
Model:IC-900 & IC-600
Overview:Ion Chromatography is widely used in environment, food, medical, chemistry, water quality, air quality detection, etc.
1. Intuitive observation with 5.0-inch high-brightness LCD, real-time display of parameters
2. Two-pole conductivity detector with temperature compensation and high sensitivity, suitable for laboratories with high sensitivity requirements
3. Fully compatible full PEEK flow paths are acid & alkali resistant and corrosion-free
4. Liquid level warning and self-processing system to automatically monitor low liquid level of the eluent and high liquid level of the waste, automatic pump stop and power off under abnormal conditions, safe and effective protection of the suppressor
5. More efficient patented constant current source device for current adjustment according to different conditions
6. Self-regenerating membrane suppressor with low background and high capacity, a perfect substitute for imported suppressors
7. “One button off” function
8. Built-in eluent preheating system to quickly and efficiently balance the eluent system
9. Imported manual six-port valve, leak proof up to 100,000 times of repeated switches
10. Self-innovated embedded column heater with imported grade A+ platinum resistance and more accurate temperature measurement
Specification | IC-900 | IC-600 |
Suppressor | ||
Type | Anion and cation, electrodialytical membrane suppression | |
Dead Volume | <90 μL | |
Background | <1.0 μS/cm | |
Capacity | >100 μeq/min | |
Injection Valve | ||
Type | 6-port, built-in, manual | |
Withstand Pressure | 35 MPa | |
Connection Pipe | 1/16” | |
Flow Path Diameter | 0.5 mm(0.02”) | |
Screw Specification | No.10-32UNF | |
Sample Loop | 25 μL | |
Pump | ||
Pump Withstand Pressure | 42 MPa (6000 psi) | 35 MPa (5000 psi) |
Max. Operating Pressure | 5000 psi | 4000psi |
Flow Range | 0.001-9.999 mL/min | |
Flow Accuracy | ≤0.1% | |
Flow Repeatability | ≤0.01% | |
Detector | ||
Type | Electrical conductivity detector | |
Material | 316 stainless steel, needle electrode | five-pole pulse ring passivated, 316 stainless steel |
Cell Volume | ≤0.8 μL | ≤1.0 μL |
Temperature Range | room temperature +5 °C-60 °C | --- |
Temperature Accuracy | ±0.01 °C | --- |
Baseline Noise | ≤0.2% FS | |
Baseline Drift: | ≤1% FS | ≤2% FS |
Quantitative Repeatability | RSD≤1.0% | RSD≤2.0% |
Qualitative Repeatability | RSD≤0.2% | |
Detection Limit | Cl-≤0.0006 μg/mL | Cl-≤0.001 μg/mL |
Column Heater | ||
Type | Imported grade A+ platinum resistance | --- |
Temperature Range | Ambient +5 °C-60 °C | --- |
Temperature Accuracy | ±0.01 °C | --- |