HPLC 5100

Category:Chromatography

Model:5100

Overview:5100 is a high-end HPLC system with independent intellectual property rights. It can load variety of configurations, such as manual valve, automatic sampler, UV detector, DAD, LIFD, RI, ELSD, etc. It is widely used in food, medicine, environmental protection, research and development.


Contact us         Inquiry

Product Features

System Configurations

Isocratic system

High-pressure binary gradient system

Quaternary low-pressure gradient system


Features

High pressure constant flow pump with parallel pump heads and optimized 5mm stroke of plunger rod ensuring perfect pump performance

Specially designed column oven for GPC represents greater capacity of three ID 8.0x300mm columns

Adopt advanced UDP communication protocol that is fast, stable, and can be used for remote control

Workstation is fully compliant to FDA 21CFR Part 11


Technical Parameters

Pump

Flow Range                                           0.001-10.000mL/min (step: 0.001mL/min)

Accuracy                                                ±0.2% (1.000mL/min, 8.5MPa, water, ambient temperature)

Precision                                                ≤0.06%RSD (1.000mL/min, 8.5MPa, water, ambient temperature)

Max Pressure                                         63MPa (0.001-2.000mL/min)

                                                               42MPa (2.001-5.000mL/min)

                                                               21MPa (5.001-10.000mL/min)

Pressure Pulsation                                  ≤1.0%


UV/Vis Detector

Wavelength Range                                190-800nm

Precision                                                 ≤0.1nm

Accuracy                                                 ≤±1.0nm

Optical Bandwidth                                   6nm

Rise Time                                                0.1-5s

Linear                                                      ≥2.0AU (5%)

Noise                                                       ≤±0.25x10-5AU

Drift                                                          ≤0.5x10-4AU/h

Minimum Detectable Concentration        1.0x10-9g/mL (Naphthalene in Methanol Solution)

Scanning Spectrogram                            Yes

Time-wavelength Program                       Yes


Diode Array Detector

Wavelength Range                                  190-800nm(512/1024 Diodes)

Spectral Resolution                                 1.2nm/0.6nm

Accuracy                                                  ±1.0nm

Linear                                                       ≥2.0AU

Light Source                                            Deuterium lamp and Tungsten lamp

Noise                                                       ≤±1.0x10-5AU

Drift                                                          ≤2.0x10-4AU/h


Column Oven

Temperature Range                                 Ambient +5℃-85℃

Precision                                                 ≤±0.1℃

Accuracy                                                 0.1℃

Temperature Setting                               0.1 ℃