Oil Content Analyzer

Category:Others

Model:ZH-800

Overview:ZH-800 infrared spectrophotometer is mainly used in the measurement of oil content in surface water, groundwater and industrial wastewater. This instrument is a professional instrument for measuring oil content in water, gas and solid.


Contact us         Inquiry

Product Features

ZH-800 Infrared Spectrophotometer is mainly used in the measurement of oil content in surface water, groundwater and industrial wastewater. This instrument is a professional instrument for measuring oil content in water, gas and solid.

High accuracy: the error of each measurement does not exceed ±2% (40㎎/L)

Baseline drift is less than 0.005AU/60min

Electrically modulated light source to improve instrument reliability

Automatic zero adjustment, automatic adjustment of instrument full scale value

The instrument uses RS232 or USB2.0 interface, which is convenient to connect with computer

With mathematical statistics, spectrum display, storage, printing and other functions

With the function of measuring instrument correction factor

Direct reading of non-dispersive measurement results without conversion

Can test the purity of carbon tetrachloride

Other non-carbon-chlorine organic solvents such as S-316, carbon tetrachloride, tetrachloroethylene, trichlorotrifluoroethane can be used as extractants

It can be extended to field-type infrared spectrophotometer, which can be driven by 220V AC power supply or external DC power supply (it can be charged for 8 hours). After expansion, it can be applied to the measurement of oil concentration in water in the field or emergency accident site, and can also perform accurate qualitative and quantitative analysis of oil samples in laboratories using AC power.

Technical Parameters

Baseline drift is less than 0.005AU/60min

The minimum detected concentration is 0.002mg/L (oil concentration in water)

Maximum measurement concentration: 64000mg/L (oil concentration in water)

Wave number range: 3400cm-1~2400cm-1 (2940nm~4167nm)

Wavenumber accuracy: ±2cm-1

Wavenumber repeatability: ±2cm-1

Absorbance range: 0.00000~3.00000AU

Basic measuring range: 0.20~100.00mg/L;

Repeatability: relative standard deviation RSD < 2% (standard oil sample with a concentration of 20-40 mg/mL)

Linear correlation coefficient R>0.999

Methyl resolution>3%

Sample analysis time: Full spectrum scan: 30 seconds

Fixed point multi-scan: 10 seconds

Non-dispersive infrared method: 2 seconds

Blank Solvent Zero: Automatic

Dimensions of the host: 470mm (length) × 330mm (width) × 150mm (height)

Weight: 10Kg

Power supply: 220VAC, 50Hz

Operating Conditions 

Stable power supply requirements, it is recommended to connect 1-5KW electronic AC stabilized power supply

The room temperature is 5-35℃, and the relative humidity is 20%~85% (RH)

No severe vibration, no strong electromagnetic field interference, no corrosive gas
Well ventilated 